| No | Hospital | Device | Tanggal | Problem | Solusi | WH | PIC |
|---|---|---|---|---|---|---|---|
| 81 | Al Ihsan RS | 0SXA2DQA | 2026-01-06 |
Flow Alarm
|
Perbaikan flow dan degess
|
12480h | Nyangnyang |
| 82 | Al Ihsan RS | 9SXA0YB1 | 2026-01-06 |
Water Alarm
|
Bersihkan filter inlet dan filter hidrolik
|
9540h | Nyangnyang |
| 83 | Al Ihsan RS | 5SXAHC73 | 2026-01-06 |
Ciklik pht
|
Cek filter Uf dan Kalibrasi TMP
|
29389h | Nyangnyang |
| 84 | Al Islam Bandung - HD Unit | 7SXATS77 | 2026-01-06 |
Flow alarm
|
Perbaikan Flow dan degess
|
30206h | Nyangnyang |
| 85 | Al Islam Bandung - HD Unit | 8SXAZD60 | 2026-01-06 |
Bibag tidak ngisi
|
Perbaikan flow dan degess
|
30585h | Nyangnyang |
| 86 | Al Islam Bandung - HD Unit | 0SXA1RKJ | 2026-01-06 |
F03 Diasafe
|
Kalibrasi TMP
|
21658h | Nyangnyang |
| 87 | Cibitung Medika RS | 8SXA0AQ3 | 2026-01-06 |
Cyclic pht f01
|
Adjsut spring uf
Pembersihan seluruh filter |
27188h | Firza |
| 88 | RS Hermina Grand Wisata | 2SXA4A3S | 2026-01-05 |
Rinse failure F16
|
Pembersihan filter uf
Adjust spring uf |
13915h | Firza |
| 89 | RS Hermina Grand Wisata | 2SXA4A4E | 2026-01-05 |
Rinse failure F04
|
Pembersihan V84
Pembersihan filter disinfektan |
13704h | Firza |
| 90 | Wisata Holistic RS - HD Unit | 9SXA0VA4 | 2026-01-05 |
Air detected
|
setting sensor udara pada board venous di EBM (LDA1&LDA2)
|
5026h | Qiyad |
| 91 | Wisata Holistic RS - HD Unit | 9SXA0VA3 | 2026-01-05 |
Air detected
|
setting sensor udara pada board venous di EBM (LDA1&LDA2)
|
4380h | Qiyad |
| 92 | Wisata Holistic RS - HD Unit | 9SXA0VA2 | 2026-01-05 |
TMP tinggi, air detected
|
Kalibrasi TMP dan setting sensor udara pada board venous di EBM (LDA1&LDA2)
|
4823h | Qiyad |
| 93 | Al Islam Bandung - HD Unit | 12ME0066 | 2026-01-05 |
Produk turun
|
Menclean P.14.2 R. 10.1 T. 170
|
59366h | Nyangnyang |
| 94 | Al Islam Bandung - HD Unit | 0SXA1Y58 | 2026-01-05 |
Upper temp.
|
Ganti sensor temp. di Heater blok
|
19612h | Nyangnyang |
| 95 | Al Islam Bandung - HD Unit | 7SXATE17 | 2026-01-05 |
T1 Failed setelah selesai pasien
|
Ganti Diasafe
|
32702h | Nyangnyang |
| 96 | RS Pindad Bandung | 7SXASE85 | 2026-01-05 |
Flow alarm
Cyclik pht |
Ganti motor degas
Kalib tmp |
17673h | Ibnu |
| 97 | RS Pindad Bandung | 7SXASE85 | 2026-01-05 |
Flow alarm
Cyclik pht |
Ganti motor degas
Kalib tmp |
17673h | Ibnu |
| 98 | RS Bayukarta Karawang | 0SXA1YKG | 2026-01-03 |
Conductiviti turun (motor degass mati)
|
Penggantian arang motor degass dan pembersihan filter
|
12528h | Qiyad |
| 99 | Puri Asih RS | 9SXA0U3L | 2026-01-03 |
Bibag jebol
|
Perbaikan dengan di lem
|
25876h | Qiyad |
| 100 | El-Syifa RS | 5V5AZM69 | 2026-01-03 |
Upper flow alm
|
Ganti AB DGS dan FLW
|